The AEMC L261 Simple Logger II is a one channel recording device powered by an alkaline battery pack. Line tracking is performed such that 64 samples over one line cycle are taken. Frequency tracking is performed over the range of ±2Hz around the nominal line frequency (50 or 60Hz). Harmonic measurements are calculated from these 64 samples (Harmonics are only available from the Simple Logger® II Control Panel within the DataView® software).
The Simple Logger® II records TRMS at a rate of up to eight times per second. The measurement process is performed no more than eight times per second. TRMS calculations are performed on a single line cycle. This means that the input(s) are ignored between measurement intervals.
The main advantage of the logger is its ability to perform a variety of recording tasks with easy and intuitive setup from a computer using DataView® software.
Analog information on the input is sampled and converted to a digital signal. This digital signal is processed and stored along with scale and time information. An optically isolated Universal Serial Bus (USB) port provides for the transfer of data from the instrument’s internal memory to the computer for analysis.
Applications:
What's Included:
The Simple Logger® II records TRMS at a rate of up to eight times per second. The measurement process is performed no more than eight times per second. TRMS calculations are performed on a single line cycle. This means that the input(s) are ignored between measurement intervals.
The main advantage of the logger is its ability to perform a variety of recording tasks with easy and intuitive setup from a computer using DataView® software.
Analog information on the input is sampled and converted to a digital signal. This digital signal is processed and stored along with scale and time information. An optically isolated Universal Serial Bus (USB) port provides for the transfer of data from the instrument’s internal memory to the computer for analysis.
Applications:
- Surge and Sag recording
- Long term supply monitoring
- Industrial, commercial, residential monitoring
- Monitor voltage harmonics
- Find intermittent voltage problems
- Machine monitoring
What's Included:
- Set of two color-coded 5ft voltage leads and color-coded alligator clips (red/black) Cat. #2140.62
The Good:
The Bad:
The Bottom Line:
The Bad:
The Bottom Line:
Full Review
- TRMS voltage recording up to 600VAC/DC
- 64 samples per cycle
- Programmable storage rates from 8 every second to 1 every day
- 3 user selectable storage modes
- Stores up to 240,000 measurements in non-volatile memory
- Powered by standard Alkaline batteries
- Lightweight, compact, fits anywhere
- 5 LED indicators quickly and clearly display Logger status
- Includes DataView® software for data storage, real-time waveform display, analysis and report generation
- Isolated USB communication
- 300V Cat IV; 600V Cat III
| ELECTRICAL | |
| Channels | One |
| Input Connection | Two recessed 4mm safety banana jacks |
| Measurement Range | 0 to 600Vac/dc |
| Resolution | 0.1V |
| Accuracy | (50/60Hz) 0 to 5V: unspecified 5 to 50V: ± (0.5% of Reading + 1V) 50 to 600V: ± 0.5% of Reading + 0.5V) |
| Sample Rate | 64 samples/cycle |
| Storage Rate | Programmable from 125ms to 1 day |
| Storage Modes | Start/Stop, FIFO and Extended Recording Mode (XRM™) |
| Recording Length | 15 minutes to 8 weeks, programmable using DataView® |
| Memory | 240,000 measurements (512KB) - The recorded data is stored in non-volatile memory and will be retained even if the battery is low or removed. |
| Communication | USB 2.0 optically isolated |
| Power Source | 2 x 1.5V AA-cell Alkaline batteries (included) |
| Battery Life | 100 hours to >45 days (dependent on sample rate and recording length) |
| MECHANICAL | |
| Dimensions | 4.94 x 2.75 x 1.28" (125 x 70 x 32mm) |
| Weight (with battery) | 6.4 oz (181 grams) |
| Case | UL94-V0 |
| Vibration | IEC 68-2-6 (1.5mm, 10 to 55Hz) |
| Shock | IEC 68-2-27 (30G) |
| Drop | Drop IEC 68-2-32 (1m) |
The DataView data logger software configures and control the power
analysis tests of your AEMC data logger from you PC-based computer.
DataView Features
• Display and Analyze Real-Time Data on PC
• Record Real-Time to PC
• Configure all Data Logger Functions and Parameters
• Display Waveforms, Trend Graphs, Harmonics (AC Models) and Text Summaries All Real-Time
• Print Custom and Standard Reports and Templates
Requirements:
DataView Features
• Display and Analyze Real-Time Data on PC
• Record Real-Time to PC
• Configure all Data Logger Functions and Parameters
• Display Waveforms, Trend Graphs, Harmonics (AC Models) and Text Summaries All Real-Time
• Print Custom and Standard Reports and Templates
Requirements:- Windows XP /
Windows Vista / Windows 7 (32/64 bit)
- 256MB of RAM for Windows XP
- 1GB of RAM for Windows Vista & Windows 7 (32 bit)
- 2GB of RAM for Windows Vista
& Windows 7 (64 bit)
- 80MB of hard disk space (200MB recommended )
- CD-ROM drive